This additional information received on 08jun2018 as follows: pt allegedly received an implant on (b)(6) 2014 as prophylaxis.Pt is alleging tilt, vena cava perforation, device is unable to be retrieved, bleeding, pulmonary embolism and long-term anticoagulation therapy.Pt.Further alleges emotional and physical pain, chest pain, hospitalization, mesenteric (organ) perforation and dvt post-implantation.Attempted retrieval on (b)(6) 2014 was unsuccessful.
|
Exemption number: e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer.Mfr site: name and address for importer site: (b)(6).Registration no: (b)(4).Additional information-investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tilt, vc & organ perforation, unable to retrieve, bleeding, pe, pain, dvt, anticoagulation '.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Unknown if the reported pain/discomfort is directly related to the filter.Unknown if the reported bleeding, anticoagulation is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Unknown if the reported peripheral vein thrombosis is directly related to the filter and unable to identify a corresponding failure mode at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|