This report is being submitted due to a retrospective review conducted under capa(b)(4).(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).Device evaluation by manufacturer: on (b)(6) 2016, an fse conducted follow-up with the customer over-the-phone to address the reported event.Fse performed troubleshooting with the customer and found that after the customer had a pm done friday, they are now getting xy errors and the needle is bent.The fse sent a sample needle to the customer.The customer replaced the needle and the unit is operational.The most probable cause of the reported event was due to a bent sample needle.
|
On (b)(6) 2016, a customer reported xy errors and needle bent with their g8 analyzer.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|