• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE ACUITY; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE ACUITY; IMPLANTABLE LEAD Back to Search Results
Model Number 4675
Device Problems Device Contamination With Biological Material (2908); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fever (1858); Post Operative Wound Infection (2446)
Event Date 03/15/2018
Event Type  Injury  
Manufacturer Narrative
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Event Description
Boston scientific received information that the left ventricular (lv) lead was part of system revision due to infection.Reportedly, the patient had fever post implant and was admitted due to wound infection.The patient was given antibiotics.It was also reported that this lv lead had contamination in the electrode end.This lv lead was explanted.There were no additional adverse patient effects reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, visual inspection of the lead was performed.Inspection of the lead tip area found a cut which severed the e4 cable and a dried body fluid or blood in the lumen of the tip area was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7443643
MDR Text Key105934826
Report Number2124215-2018-06444
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/05/2019
Device Model Number4675
Other Device ID NumberACUITY X4 SPIRAL S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0295; 4470; 4675; G148
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age56 YR
-
-