ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
Final analysis found burn marks on the ac power adapter and its main circuit board. Further analysis found that after opening the transmitter, burnt modem circuit components were observed on the main pcb. It was concluded that the root-cause of the field complaint of no power issue was the burn damages noted on the ac power adapters pcb and transmitter main pcb.
 
Event Description
It was reported that the patient's transmitter would not power up. Multiple outlets were attempted and prongs were reset. The transmitter was returned and replaced. No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7444518
MDR Text Key106016496
Report Number2017865-2018-06061
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberEX1150
Device Catalogue NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-