Results: the lantern was ovalized approximately 112.0 cm from the hub to the distal tip.Conclusions: evaluation of the returned device revealed that the distal tip of the lantern was ovalized.This type of damage typically occurs due to improper handling during use.If the distal tip of the catheter is forcefully gripped or pinched during insertion, damage such as this may occur.The ovalized distal tip of the lantern likely contributed to the reported resistance and prevented the lantern from being advanced into the non-penumbra catheter.The non-penumbra catheter referred to in the complaint was not returned to penumbra for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a coil embolization procedure using a lantern delivery microcatheter (lantern).During the procedure, the physician was opened and flushed the lantern, and then attempted to advance the lantern into the non-penumbra catheter using a guidewire.However, the physician felt resistance and was unable to advance the lantern into the hub of the catheter, despite attempting twice.The lantern was therefore removed, and the procedure was completed using a new lantern.There was no report of an adverse effect to the patient.
|