MDT PUERTO RICO OPERATIONS CO SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical product: product id: 8711, lot# unknown, product type: catheter. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a foreign healthcare provider (hcp) via a manufacturer's representative regarding a patient who was receiving baclofen (4000 mcg/ml at 1580. 6 mcg/day) via an implantable pump for an unknown indication for use. It was reported increased aspirated volumes were observed during the last three refills. First, the difference between actual reservoir volume (arv) and expected reservoir volume (erv) was 4 ml. The refill after, the difference was 5 ml, and at the last refill the week of (b)(6) 2018, the difference was 6 ml. It was stated the catheter was currently implanted for about 23 years, according to the hcp. It was reported the pump logs did not show any anomalies. There were no reported symptoms. No complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7445280
MDR Text Key106031991
Report Number3004209178-2018-08444
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 04/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/14/2015
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/26/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured11/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-