COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Complete Blockage (1094); Inadequate Filtration Process (2308); Obstruction of Flow (2423); Structural Problem (2506); Unintended Movement (3026); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Pulmonary Embolism (1498); Dyspnea (1816); Internal Organ Perforation (1987); Perforation of Vessels (2135); Stenosis (2263); No Information (3190); Thrombosis/Thrombus (4440)
Event Type  Injury  
Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It is alleged that the plaintiff received a gunther filter on (b)(6) 2007.It is alleged that the plaintiff was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 10may2018 as follows: ¿[pt] allegedly received an implant on (b)(6) 2007 via the right common femoral vein due to deep vein thrombosis, pulmonary embolism.[pt] is alleging tilt, organ perforation.[pt] further alleges filter clogged up in (b)(6) of 2004 and 2005, causing clots, had to go on clog buster (shots in the stomach 2 times a day) and now on xarelto (blood thinners), get out of breath easily.
 
Manufacturer Narrative
Additional information: investigation.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿gunther tulip filter implanted".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Additional information: investigation.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿gunther tulip filter implanted".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
[pt] further alleges filter clogged up in (b)(6) of 2014 causing clots, had to go on clog buster (shots in the stomach 2 times a day) and now on xarelto (blood thinners), get out of breath easily.
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tulip -thrombosed ivc, tilt, organ perforation, blood clots (potential pe), shortness of breath'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported shortness of breath is directly related to the filter.Rpn and lot# are unknown, but the tulip filter is manufactured and inspected according to "specifications".No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: investigation.Correction: b7.The following fields were updated per additional information received: b5, b7, h6.H6 conclusion code(s): appropriate term/code not available (4316) was selected because the previous investigation remains unchanged by the additional information.Investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Additional information has been provided.A percutaneous routine retrieval was attempted (b)(6) 2014.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: investigation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation, stenosis.The additional information regarding vena cava perforation does not change the previous investigation results for organ perforation.The additional information regarding stenosis does not change the previous investigation results for inferior vena cava (ivc)/filter thrombosis.The catalog # and lot # are unknown, but the filter is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Report from ct (computed tomography): "filter type: aln.Ivc stenosis: yes.Filter position: below the level of the renal veins.Impressions: the ivc is collapsed around the filter which is consistent with chronic ivc stenosis.Axial image 35.The ivc distal to the filter is stenotic.The anterior strut penetrates 3.5 mm through the ivc wall.Axial image 39.The left strut penetrates 4.1 mm through the ivc wall.Axial image 39.The posterior strut penetrates 4.5 mm through the ivc wall.Axial image 39.The right strut penetrates 4.1 mm through the ivc wall.Axial image 39.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key7450378
MDR Text Key106157396
Report Number1820334-2018-01161
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/18/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient SexFemale
Patient Weight84 KG
-
-