• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Device Inoperable (1663); Defective Component (2292)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported, via the investigation results, that the mixing pump had failed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7453661
MDR Text Key106411864
Report Number1018233-2018-01351
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-