MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOMECHOICE CLARIA; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOMECHOICE CLARIA; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C6M10
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice claria under delivered during peritoneal dialysis therapy.This issue occurred during the last fill while the patient was connected.The patient stated that the homechoice device completed treatment during the last fill, however the next drain on the following day the patient could not remove any fluid.The patient stopped the treatment and visited the hospital to find out that the patient was empty and the machine had not delivered the fluid it was supposed to.There was no other information reported.The renal therapy service agent discussed the use of continuous ambulatory peritoneal dialysis.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.External/internal visual inspection with naked eye was performed with no issues noted.Full functional testing, electrical safety testing, calibration and simulated therapy was performed with no issues noted.The reported issue was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE CLARIA
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7454611
MDR Text Key106292746
Report Number1416980-2018-02342
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C6M10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/29/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-