MEDTRONIC EUROPE SARL PERCEPTA CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W1TR04
Device Problems Failure to Interrogate (1332); Pacing Problem (1439)
Patient Problem Syncope (1610)
Event Date 04/12/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced syncope.The device could not be interrogated and the electrogram showed no pacing spikes (output.) the device was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Battery analysis revealed that the anode insulation system was breached, creating an internal short condition.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary : the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Battery analysis revealed that the anode insulation system was breached, creating an internal short condition.Hybrid analysis discovered dendrite under the c12 ceramic capacitor causing current leakage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA CRTP
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key7461597
MDR Text Key106599957
Report Number9614453-2018-01631
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/28/2018
Device Model NumberW1TR04
Device Catalogue NumberW1TR04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2414-2018
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age95 YR
-
-