Model Number 16-02-85
Device Problem Insufficient Heating (1287)
Patient Problem No Patient Involvement (2645)
Event Date 02/05/2018
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to confirm the reported issue.Visual inspection was performed and no visible defects were observed.A functional test was performed without issue and a temperature check-up on both the patient and cardioplegia tanks found that all temperatures were reached within normal amount of time.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.As the issue could not be reproduced, a root cause was not determined.H3 other text : evaluated on site by livanova rep.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t experienced inadequate warming on the patient side during setup.The device reportedly would not heat fast enough.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w 65th way
arvada, CO 80004
2812287260
MDR Report Key7466341
MDR Text Key106864052
Report Number9611109-2018-00935
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public010403381790052811170529
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-