ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0700
Device Problems Failure to Calibrate (2440); Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/31/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the fiber optic sensor (fos) was reconnected to the intra-aortic balloon pump (iabp) and tones were heard, but there was no signal. The pump continued to read "fos not connected". After multiple attempts to calibrate (cal) the mean arterial pressure (map), the clinical support specialist (css) had the rn unplug the fos and cal key to allow the pump to use the transducer. The pump was showing a waveform according to the rn, but still showed fos. After a second call it was determined that the pump monitor never showed fos, but the rn was reading it on the ap menu screen. The rn swapped out the pump and was able to switch consoles and the fos still was not connecting. As a result, the transducer was used without issue and the pump was giving appropriate support to the patient. There was no report of patient complication or serious injury and death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAC3 OPTIMUS IABP NA/EMEA
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
carmen sherman
16 elizabeth drive
chelmsford, MA 01824
9782505100
MDR Report Key7466573
MDR Text Key106715323
Report Number3010532612-2018-00099
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberIAP-0700
Other Device ID Number00801902084965
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-