STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 STERNUM SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 7207000000 |
Device Problems
Sticking (1597); Positioning Problem (3009)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/02/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.2 events were reported for this quarter.2 devices were received for evaluation.2 events were confirmed during testing.2 devices were found to be affected by corrosion 2 devices were repaired and returned to the account.There were no remedial actions taken.This device is not labeled for single-use.
|
|
Event Description
|
This report summarizes 12 malfunction events in which the device had a stuck safety switch, which can cause unintended activation.There was no patient involvement; no patient impact.
|
|