• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 STERNUM SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 STERNUM SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 7207000000
Device Problems Sticking (1597); Positioning Problem (3009)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2017
Event Type  malfunction  
Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.2 events were reported for this quarter.2 devices were received for evaluation.2 events were confirmed during testing.2 devices were found to be affected by corrosion 2 devices were repaired and returned to the account.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 12 malfunction events in which the device had a stuck safety switch, which can cause unintended activation.There was no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 7 STERNUM SAW
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zachary baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key7466627
MDR Text Key106956023
Report Number0001811755-2018-00677
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported12
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number7207000000
Was Device Available for Evaluation? No
Date Manufacturer Received12/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-