Model Number ORCHESTRA PLUS |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Unexpected Shutdown (4019)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/04/2018 |
Event Type
malfunction
|
Event Description
|
Reportedly, the subject programmer suddenly shutdown and reboot in loop.Preliminary analysis of the returned programmer revealed that the reported issue could be linked to a real time clock (rtc) issue.
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Event Description
|
Reportedly, the subject programmer suddenly shutdown and reboot in loop.Preliminary analysis of the returned programmer revealed that the reported issue could be linked to a real time clock (rtc) issue.
|
|