DATASCOPE FAIRFIELD TRANS-RAY PLUS 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0605
Device Problem Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2018
Event Type  malfunction  
Manufacturer Narrative
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.(b)(4).
 
Event Description
It was reported that in an emergent case during intra-aortic balloon (iab) insertion in an angina pectoris patient the iab was unable to advance through the sheath.The iab was replaced to continue therapy.Mild sclerosis, calcification and tortuosity were noted in the patient vessel.No patient injury was reported.
 
Event Description
It was reported that in an emergent case during intra-aortic balloon (iab) insertion in an angina pectoris patient the iab was unable to advance through the sheath.The iab was replaced to continue therapy.Mild sclerosis, calcification and tortuosity were noted in the patient vessel.No patient injury was reported.
 
Manufacturer Narrative
The iab was returned with the membrane loosely folded.No blood was visible on the catheter.A non-maquet guide wire was also returned partially inside the iab lumen.The sheath was not returned for evaluation.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.An evaluation of the product was unable to duplicate the reported problem.The product performed according to specification.A device and lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Complaint # (b)(4), record # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key7471649
MDR Text Key106978936
Report Number2248146-2018-00286
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2020
Device Catalogue Number0684-00-0605
Device Lot Number3000059179
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Device AgeYR
Date Manufacturer Received05/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-