• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808012300
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter in two pieces.The balloon was loosely folded.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube was completely separated 68.6cm from the hub.The fracture faces were oval as if kinked prior to separation.There are numerous hypotube kinks.The tip is damaged.There was no evidence of any material or manufacturing deficiencies contributing to the damage. the investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
Reportable based on device analysis completed on 06-apr-2018.It was reported that shaft kink occurred.The target lesion was located in a tortuous and stenosed coronary artery.A 3.0mm x 12mm quantum¿ maverick¿ balloon catheter was advanced to dilate the lesion.However, it was noted that the device delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a hypotube break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7476770
MDR Text Key107039362
Report Number2134265-2018-03560
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392675
UDI-Public08714729392675
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/13/2020
Device Model NumberH7493808012300
Device Catalogue Number38080-1230
Device Lot Number21136653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-