Model Number N/A |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Code Available (3191)
|
Event Date 02/15/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).
Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.
Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that while attempting to move the set screw the set screw became stuck in the nail, and the product could not be used.
The set screw was used to move the lock down on the a.
R.
And log screws.
However, this could not be done as the set screw became stuck in the nail while moving it with the hex driver.
Subsequently, the nail was removed.
Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.
Device history record was reviewed and no discrepancies relevant to the reported event were found.
Root cause was unable to be determined.
A summary of the investigation has been sent to the complainant.
If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|