• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The patient experienced adverse events on different days with the same device.The following reports are being submitted: (b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the receiver initialized without a manual restart.No additional event or patient information is available.No product or data were provided for evaluation.The complaint confirmation of the receiver initializing without manual restart could not be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint receiver device was returned for evaluation.The device was externally visually inspected and no defects were found.The receiver was charged and rebooted.The receiver log was downloaded, there was no data within the investigation window.Functional testing was performed and it passed.The reported event of receiver showing initializing screen without manual restart could not be determined.The root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key7481798
MDR Text Key107194186
Report Number3004753838-2018-47058
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-1
Device Catalogue NumberSTK-GL-010
Device Lot Number5231317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
-
-