Model Number 4415
Device Problems Complete Blockage (1094); Insufficient Cooling (1130)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.The field service representative (fsr) duplicated the reported complaint.The cpg water would not go below 12 degrees celsius.The large tank and small tank filters were clogged.The filters were cleaned and the water reached seven degrees celsius.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that prior to use of the device for a cardiopulmonary bypass (cpb) procedure, the cardioplegia (cpg) water temperature would not get cold enough.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7486644
MDR Text Key107568183
Report Number1828100-2018-00238
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-