Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
According to the available information, tubing break from cylinder to pump.Additional information received indicated the ipp was originally placed in 2005.The malfunction was noted about 2016.The device was "working good" for about 10 years.The patient believes that his penile length is shorter post-ipp than before.In about 2016, the device stopped working, the patient would press the pump as usual, but the cylinders would not inflate.The doctor stated, upon pressing the pump, cylinders do not inflate, pump stays deflated (c/w leak).In the or, the proximal portion of the tubing on the patient's right cylinder appeared to be the source of leakage, with a partial break in the tubing.
|