• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS REFURB, SR PREVUE PLUS; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS REFURB, SR PREVUE PLUS; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770086R
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
It was reported by the biomedical engineer that the active ultrasound image is displayed on 2/3 of the display but the whole screen displays artifact in the form of lines running through it.
 
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of only 2/3 of the screen works and it has lines through it was confirmed.The reported issue of only 2/3 of the prevue lcd screen works and it has lines through it is confirmed.The root cause is determined to be due to a puncture in the front enclosure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to refurbished inventory.
 
Event Description
It was reported by the biomedical engineer that the active ultrasound image is displayed on 2/3 of the display but the whole screen displays artifact in the form of lines running through it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SR PREVUE PLUS
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7492884
MDR Text Key107695500
Report Number3006260740-2018-00909
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741124945
UDI-Public(01)00801741124945
Combination Product (y/n)N
PMA/PMN Number
K150529
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770086R
Device Catalogue Number9770086R
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Event Location Hospital
Date Manufacturer Received03/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-