PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Occlusion (1984)
Event Date 03/06/2018
Event Type  Injury  
Manufacturer Narrative
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Re-occlusion (distal embolization) is a known and anticipated complication with these types of procedures and are noted in the device labeling.Therefore, it was determined that the reported event was an anticipated procedural complication.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.It is unknown why the device has not been returned for evaluation.
 
Event Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).It was reported that there was an occlusion of the frontal branch of the middle cerebral artery, which arose from the neck of the aneurysm.The physician retracted and removed the smart coil, and performed additional treatment to address this device related adverse event, and the event is considered resolved.The occlusion was adjudicated to be a device related adverse event with a definite relationship to the smart coil system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7494933
MDR Text Key107631795
Report Number3005168196-2018-00947
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age48 YR
-
-