The report of a kinked outflow graft was confirmed via review of a submitted ct scan image while the report of low flow alarms was confirmed through the evaluation of the submitted log files and log files downloaded from the returned device; however, no device-related issues were discovered during the evaluation of heartmate 3 lvad.Review of the submitted controller event log file revealed continuous low flow hazard alarms throughout the entire log file.The submitted and downloaded lvad periodic log file data showed that the low flow events began as early as (b)(6) 2018.Despite the low flow condition, the pump was able to operate at the set speed without issue.The returned portion of the sealed outflow graft showed no evidence of distortion such as twisting or kinking.The lumen of the graft attachment and the outflow graft material revealed only traces of post-explant blood with no evidence of any adhered or developed depositions or thrombus formations.Visual examination of the remaining blood-contacting surfaces upon disassembly of the returned pump revealed a small amount of post-explant blood with no evidence of any adhered or developed depositions or thrombus formations.Review of the submitted ct scan image showed that the outflow graft was kinked in the area not covered by the bend relief.Upon evaluation of the returned device, it appeared that the center had removed the portion of the sealed outflow graft which contained the kink and anastomosed a new portion of graft material in its place.The kink in the outflow graft confirmed through the submitted ct image could have potentially contributed to the low flow events captured in the log file data.Visual inspection of the pump rotor and rotor well did not reveal any surface scratches or defects.The pump was cleaned, reassembled, and operated on a mock circulatory loop.The device functioned as intended and in accordance with manufacturing specifications.The heartmate 3 lvas instructions for use (ifu) contains information on preparing the sealed outflow graft and instructs the user to verify that the graft is not twisted or kinked by checking the position of the black line on the graft above and below the bend relief and ensuring that the line is straight.In addition, the ifu outlines all pump parameters, provides information regarding the assessment of pump flow, and outlines all system controller alarms and the appropriate actions associated with each alarm condition.A review of the device history records, including the sterilization and packaging documentation, found no deviations from manufacturing specifications.The manufacturer is closing the file on this event.
|