MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 7 ELITE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 7 ELITE; SCS IPG Back to Search Results
Model Number 3662
Device Problem Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/11/2018
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported that the physician used electrocautery to stop bleeding at the wound check appointment.The implanted ipg was unable to connect with the external devices.Reportedly, the ipg was not placed into surgery mode and deemed inoperable.As a result, surgical intervention was undertaken wherein the ipg was explanted and replaced.Patient was successfully programmed.The issue was resolved with ipg replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 7 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7498750
MDR Text Key107735063
Report Number1627487-2018-04308
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/23/2019
Device Model Number3662
Device Lot Number6164870
Other Device ID Number05415067020222
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Treatment
MODEL 1192 (2), SCS ANCHOR; MODEL 3186(2), SCS LEAD
Patient Outcome(s) Other;
Patient Age85 YR
-
-