Model Number 50000000
Device Problems Air Leak (1008); Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device was not returned.
 
Event Description
It was reported that the complainant needed assistance with filling the device. The water level read two bars and the device was not alarming low reservoir; however, the device was displaying air leak and the flow rate was 0. 0 l/min. It was noted that the pads were exchanged prior to troubleshooting and the flow rate was still 0. 0 l/min. Per troubleshooting, the fluid delivery line was removed and there was no suction with finger over port, the flow rate was still 0. 0 l/min. The complainant stated they were unaware if there was another device in their facility. It was advised that the device be sent to biomed to be evaluated for low flow. Per additional information received the patient was placed on a cooling blanket and the device was sent to biomed for evaluation. Biomed stated that after troubleshooting with tech support the circulation pump needed to be replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7499800
MDR Text Key107970851
Report Number1018233-2018-01578
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741080142
UDI-Public(01)00801741080142
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-