• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ELECTROSURGICAL KNIFE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ELECTROSURGICAL KNIFE Back to Search Results
Model Number KD-611L
Device Problem Device Handling Problem (3265)
Patient Problem Perforation (2001)
Event Date 04/17/2018
Event Type  Injury  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc).The exact cause has been under investigation.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
During an endoscopic submucosal dissection, the subject device was used and perforation occurred.The physician closed the perforation with clips and the procedure was completed.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.This type of the event is most likely related to the operator's technique.Based on the past similar cases, it was known that the event occurred due to any of the following possible causes.The high-frequency output was set too high.The activation time was too long.The output was activated when in contact with adhering fluids.Incised the tissue by pressing the knife forcefully more than necessary.The instruction manual of the device has already warned as follows; do not set the output value of the electrosurgical unit too high or too low.Also, do not allow the activation time to be too long or too short.Set the high-frequency output mode of the electrosurgical unit optimally according to the conditions of the tissue to be cut.An excessive or insufficient output value may result in perforation, bleeding, mucous membrane damage or thermal injuries to the non-target tissue.Aspirate fluids such as mucus that adhere to the electrode and/or the cutting knife, outer sheath and body cavity tissues.Patient injury such as punctures, hemorrhages, mucous membrane damage and thermal injury of tissue could result if output is activated when in contact with these adhering fluids.When current is discharged while the cutting knife is being separated from the mucosa under wet situation, it may break the cutting knife or crack the tip.This instrument can be incised by both the cutting knife and the electrode.When incising the tissue by pressing the knife forcefully more than necessary, unintended incision may occur by the electrode part, and it may cause perforation and massive bleeding.Always confirm the direction of incision and do not incise with more pressure than necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE
Type of Device
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key7501285
MDR Text Key107821458
Report Number8010047-2018-00797
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
PMA/PMN Number
K092309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberKD-611L
Device Lot Number82K
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-