DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problem
Device Operational Issue (2914)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the small battery drive device stopped turning.It was not reported if there were any delays to the surgical procedure or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.It was determined that the motor seized, jammed and was moving heavily.It was further determined that the device had no function.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|