BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-1132
Device Problems Migration or Expulsion of Device (1395); Charging Problem (2892)
Patient Problems Pain (1994); Respiratory Distress (2045)
Event Date 01/30/2018
Event Type  Injury  
Manufacturer Narrative
Additional suspect medical device components involved in the event: model#: sc-2317-70, serial#: (b)(4); description: infinion cx 70 cm lead.
 
Event Description
A report was received that the patient was experiencing pain in the ribs and was having difficulty breathing due to the charging belt that had to be put on very tightly when charging the ipg.It was noted that the patient was having difficulty charging as the ipg had flipped sideways.The patient will undergo an explant procedure.
 
Manufacturer Narrative
Additional information was received that no further course of action will be taken at this time.It is indicated that the devices will not be returned for evaluation; therefore a failure of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
 
Event Description
A report was received that the patient was experiencing pain in the ribs and was having difficulty breathing due to the charging belt that had to be put on very tightly when charging the ipg.It was noted that the patient was having difficulty charging as the ipg had flipped sideways.The patient will undergo an explant procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key7504336
MDR Text Key107921991
Report Number3006630150-2018-01621
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729821526
UDI-Public08714729821526
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/14/2018
Device Model NumberSC-1132
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
-
-