ARGON MEDICAL DEVICES INC. FIRST PICC 26GA, SINGLE LUMEN Back to Search Results
Catalog Number 384221
Device Problem Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 02/15/2018
Event Type  malfunction  
Manufacturer Narrative
There was no lot number available; therefore, a review of the lot number could not be conducted.There were no samples returned for review; however, an image was provided.Based on the image provided, breakage is confirmed just below the inverted triangle of the extension set.The image does not provide a clear view of the breakage site to verify a straight or jagged edge.A straight edge would indicate the catheter might have been cut and a jagged edge would be indicative of tensile force being applied to the catheter.Without the sample to review, a definite root cause and corrective action cannot be established.If the sample is returned for evaluation, a follow-up report will be submitted.
 
Event Description
¿bedside rn went to bedside @ 2200 to do hourly picc line check.She has visualized the line previously at 2000 and 2100 and it was wnl.Ar 2200, she found the picc line lying in the bed beside the baby.The line snapped at the side of the heart closest to the baby.The remaining portion of the line was still secured to the baby¿s arm by picc dressing.¿ per the photo-(b)(4):the injection cap was placed directly on the hub of picc no lot# provided.
 
Manufacturer Narrative
There was no lot number available; therefore, a review of the lot number could not be conducted.One sample was returned for evaluation.Breakage was observed just under the inverted triangle of the silicone extension set.The breakage was examined under magnification and it was noted the edges were slightly jagged which is indicative of a tensile force being applied to the catheter.Possible causes for the breakage at the inverted triangle of the extension set could be related to patient movement or securement issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRST PICC 26GA, SINGLE LUMEN
Type of Device
FIRST PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES INC.
1445 flat creek road
athens TX 75751
Manufacturer (Section G)
ARGON MEDICAL DEVICES INC.
Manufacturer Contact
gail smith
1445 flat creek road
athens, TX 75751
2144368995
MDR Report Key7508426
MDR Text Key108370411
Report Number1625425-2018-00046
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972262
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number384221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-