Catalog Number 3L92512
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Mix up of 2 lots of corail stems reported: during a product investigation, the supplier discovered that a product etched with product code 3l92512/ lot 5308136 was found inside a packaging labelled "product code 3l92512/ lot 5307580".
 
Manufacturer Narrative
Product complaint # : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).The reported event has been evaluated and confirmed.A worldwide recall was initiated for this issue.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 STD SIZE 12
Type of Device
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS 3003895575
7 allée irène joliot curie
bp 256
saint priest cedex 69801
FR  69801
MDR Report Key7510105
MDR Text Key108096727
Report Number1818910-2018-59454
Device Sequence Number1
Product Code KWA
UDI-Device Identifier10603295168782
UDI-Public10603295168782
Combination Product (y/n)N
PMA/PMN Number
K042992
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3L92512
Device Lot Number5307580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Date Manufacturer Received11/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-