MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE SCS IPG Back to Search Results
Model Number 3660
Device Problem Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/19/2018
Event Type  Injury  
Manufacturer Narrative
Concomitant information: model 3186 (2), scs lead; implant date is unknown. The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017. The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported. The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported ((b)(6)) during an unrelated surgical procedure on (b)(6) 2018, the patient's programmer reflected the error message: "generator could not be restored". It was noted the patient's ipg was turned off prior to the procedure. However, following the procedure, the device could not be turned back on. No other interventions have been planned at this time.
 
Event Description
Follow-up information revealed the patient underwent surgical intervention wherein the ipg was explanted and replaced. Reportedly, effective therapy resumed following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 5 ELITE
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7514415
MDR Text Key108262116
Report Number1627487-2018-04804
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/01/2019
Device Model Number3660
Device Lot Number6079319
Other Device ID Number05415067020192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-060217-001-C

Patient Treatment Data
Date Received: 05/15/2018 Patient Sequence Number: 1
-
-