The customer reported that a malfunction of the intercom is occurring when receiving audio on the operator side.This occurred on a brilliance ict sp system.There was no report of harm as a result of this issue.A philips field service engineer (fse) confirmed there was no report any patient impact or rescan as a result of this issue as this was discovered during a service visit.The fse evaluated the system and determined that the front and rear breathing light assemblies needed to be replaced, and confirmed that after they were replaced it was possible to hear the voice from both front and rear patient positions.The system is operational and in clinical use.This event has been determined not to be a reportable event.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|