• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported that the gantry microphone was not working and the patient could not be heard.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported that a malfunction of the intercom is occurring when receiving audio on the operator side.This occurred on a brilliance ict sp system.There was no report of harm as a result of this issue.A philips field service engineer (fse) confirmed there was no report any patient impact or rescan as a result of this issue as this was discovered during a service visit.The fse evaluated the system and determined that the front and rear breathing light assemblies needed to be replaced, and confirmed that after they were replaced it was possible to hear the voice from both front and rear patient positions.The system is operational and in clinical use.This event has been determined not to be a reportable event.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7514851
MDR Text Key108498952
Report Number1525965-2018-00400
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-