Model Number 16-02-80
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Date 04/03/2018
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The event occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and identified that the 230v pin at the power plug was burnt. The power cable was exchanged with the cable from a decommissioned device. The damaged cable was discarded by the facility and could not be investigated further. The cable connection was tested and an electrical safety test was performed. No further issues were identified and the device was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report. Discarded; not available for return.

 
Event Description

Livanova (b)(4) received a report that the plug for the internal pe-line of a heater-cooler system 3t was found to be loose during priming. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w 65th way
arvada, CO 80004
2812287260
MDR Report Key7516625
MDR Text Key108494847
Report Number9611109-2018-00960
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 05/15/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received05/15/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/03/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured01/12/2011
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-