Results: the neuron max was fractured approximately 92.0 cm from the hub.Conclusions: evaluation of the neuron max revealed that the distal tip was fractured.This damage typically occurs due to improper handling during removal from the packaging.If the packaging mandrel slips out from the packaging tape, and the neuron max and packaging mandrel are forcefully withdrawn through the packaging tube, damage such as this may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a medical procedure, the hospital staff noticed that the tip of a neuron max 6f 088 long sheath (neuron max) was ripped upon removal from the packaging.The damage to the neuron max was found prior to use.Therefore, the neuron max was not used in the procedure.The procedure was completed using a new neuron max.There was no report of an adverse effect to the patient.
|