Model Number 4028310
Device Problems Detachment Of Device Component (1104); Component Falling (1105); Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2018
Event Type  Malfunction  
Manufacturer Narrative

A trumpf medical sales rep found that it was possible to release the handle adapter without pressing the unlocking pin. The locking pin was adjusted and the handle adapter worked properly. The lamp head was replaced by trumpf service and returned for further analysis. A follow-up report will be submitted if new information becomes available.

 
Event Description

The sterile handle & adapter fell off lamp head. No injury occurred.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameILED TRIO
Type of DeviceSURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM 07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key7517758
MDR Text Key108614073
Report Number9681407-2018-00024
Device Sequence Number1
Product Code FSY
Combination Product (Y/N)N
Reporter Country CodeGM
PMA/PMN NumberK061317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Remedial Action Repair
Type of Report Initial,Followup
Report Date 04/17/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received05/16/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number4028310
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/17/2018
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured03/14/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-