• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS REFURB, SR PREVUE PLUS SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS REFURB, SR PREVUE PLUS SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770086R
Device Problem No Display/Image (1183)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device has not been returned to the manufacturer, at this time, for evaluation. A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Customer reported screen will turn dark and then go black.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s) sample(if available) applicable fmea documents, labeling and applicable manufacture records. Based on a review of this information, the following was concluded: the device was not returned for evaluation/repair. During evaluation, the reported issue of scanner displays a dark screen was confirmed. During the evaluation of the scanner it was determined that an internal failure in the beam former is the cause of the failure, leading to an unresponsive scanner. No other functionality issues with the equipment were found during evaluation/servicing. The device was serviced, tested and refurbished inventory. A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number. (b)(4).
 
Event Description
Customer reported screen will turn dark and then go black.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREFURB, SR PREVUE PLUS
Type of DeviceSYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP.
141 zehner school road
zelienople PA 16063
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7522547
MDR Text Key108687085
Report Number3006260740-2018-00994
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741124945
UDI-Public(01)00801741124945
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120822
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/16/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9770086R
Device Catalogue Number9770086R
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location Hospital
Date Manufacturer Received07/22/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-