Customer stated the analyzer had emitted smoke from the battery compartment.Upon removal of the batteries from the battery compartment the analyzer performed as expected with the a/c adapter.Information related to proper battery placement was not obtained at the time of reporting, however during the investigative process it was noted after battery insertion the analyzer would not power on.Quality control testing was conducted using an ac adaptor and the analyzer performed as expected.The sensor pins were examined during the investigation, and no corrosion was noted.There was no report from the customer of impact to patient results.Customer has received a new analyzer and has been trained on proper use of batteries and the a/c adapter for the leadcare ii analyzer.(b)(4).
|