ZIMMER MANUFACTURING B.V. COCR FEMORAL HEAD; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Detachment Of Device Component (1104)
Patient Problem No Information (3190)
Event Date 10/03/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 00-8114-000-00, cpt 12/14 size 0 cocr std, 63683211.Report source, foreign - the event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a left hip revision procedure the head disassociated from the femoral stem taper.The locking ring and femoral head were removed and other components were used to complete the procedure, resulting in a 45 minute delay.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Visual inspection of the femoral head shows some scratching inside the taper and on the etching.No other visible damage was noted.Dimensional analysis of the head was conforming to print specifications.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR FEMORAL HEAD
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7525678
MDR Text Key108632290
Report Number0002648920-2018-00151
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
PK953337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00801803202
Device Lot Number62884892
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-