• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97714
Device Problems Energy Output Problem (1431); Unable to Obtain Readings (1516); Device Operates Differently Than Expected (2913); Insufficient Information (3190); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Therapeutic Effects, Unexpected (2099); Tingling (2171)
Event Date 04/18/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 97714, serial (b)(4), implanted: (b)(6) 2017, product type: implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient via manufacturer representative regarding patient's neurostimulator (ins) that was implanted for spinal pain and non-malignant pain.It was reported that patient feels pulsating sensation in his ins pocket area when he lies on that area and when he bends over.The pulsating is like a heartbeat.Patient has been using hd programming and his current hd settings for a long time.Patient usually doesn't feel stim with his hd settings, though sometimes he can feel a light tingling.Caller tried programming ld settings today but this didn't resolve the issue.Patient does not feel sensation with stim off.Palpation of ins area with stim on did not create any changes in stim.Patient does note that when the pulsating is occurring that his stimulation does feel different.Impedances taken with patient in position that causes this pulsating did not reflect any changes in impedances.(initial check of imped did show "xxx" on all pairs with 0 except #13, which showed 2950 ohms).Caller remeasured impedances and then these values normalized.No trauma/falls were reported that could be related to this issue.Patient can turn the stim down and then this pulsating isn't felt as strongly, but then patient doesn't get the pain relief he needs when using lower voltage.The pulsating issue started in (b)(6) 2018; an exact date was not provided.They will keep an eye on the issue for now.No further complications were reported/anticipated.
 
Manufacturer Narrative
Product id: 97714, serial# (b)(4), implanted: (b)(6) 2017, product type: implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer¿s representative (rep).The rep reported that the cause of the pulsating sensation wasn¿t determined.The rep reported that the patient was trying different programming.The pulsating sensation wasn¿t resolved.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the rep.Caller stated that after initial troubleshooting and patient had time to experience updated programming.Caller stated no changes as a result and requested advice on next steps and confirmed that patient could not turn up stimulation to get pain relief.Caller was advised to run additional impedance tests in different positions, reprogram stimulation and possibly obtain imaging to look for damaged leads and discuss next steps with the managing physician.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Concomitant medical products: product id 97714, serial# (b)(4), implanted: (b)(6) 2017, product type implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer's representative (rep).The rep reported that the issue was resolved.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7526312
MDR Text Key109232777
Report Number3004209178-2018-11286
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age49 YR
-
-