LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the heater-cooler system 3t. The event occurred in (b)(6). The field service representative who identified the issue found the patient circuit pump 1 to be circulating very slowly. The service representative removed the patient circuit bridge and replaced patient circuit pump 1 to resolve the issue. A blind plug was installed in the upper overflow port and the device was cleaned under the patient bridge and in the patient tank. The cleaning procedure was completed and subsequent functional verification testing did not identify further issues. The unit was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report. Device not returned.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error code on the patient side during a cleaning cycle and the patient water circuits could not be circulated. This issue was discovered by a livanova field service representative during maintenance. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7527178
MDR Text Key108732841
Report Number9611109-2018-00975
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received08/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-