Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 04/27/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The third party service agent evaluated the device and verified the reported issue.The service agent replaced the user interface pcb assembly and the user interface to stack flex cable assembly.Thereafter, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.The device was not returned to physio-control for evaluation.
 
Event Description
A third party service agent contacted physio-control to report that their customer¿s device intermittently displayed a white screen with vertical lines.In this state the device may be inoperable and defibrillation therapy may not be available if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7527508
MDR Text Key108728636
Report Number0003015876-2018-00796
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000037
Was Device Available for Evaluation? No
Date Manufacturer Received04/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No