LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Energy Output To Patient Tissue Incorrect (1209)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2018
Event Type  malfunction  
Event Description
The patient's device was interrogated and found to be disabled.The device was previously programmed to therapeutic level.Programming history was reviewed and it shows that the device was disabled.The device was interrogated again and settings were observed to be at 0 ma.But there was no programming event after that to enable the device by the physician.
 
Event Description
It was reported that the physician did not intentional disable and from their recollection no error message internal investigation shows that the physician likely encountered the "errant invalid parameter values " error message, which forced the programming to 0 ma.The device was programmed without the physician's intention.While the device could have been programmed back by the physician, there is not a logged attempt to reprogram it back to original settings on the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7529963
MDR Text Key109000912
Report Number1644487-2018-00822
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/26/2018
Device Model Number106
Device Lot Number203782
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
-
-