SMITH & NEPHEW, INC. FOOTSWITCH,EP-1,PEDAL STYLE,; ARTHROSCOPE Back to Search Results
Model Number 7205396S
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2018
Event Type  malfunction  
Event Description
It was reported that the footswitch was sticking, causing that the shaver blade was running continuously.No patient injuries were reported.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of stuck reverse (left) pedal was confirmed.Footswitch failed functional testing using a known good control unit and mdu.The actuator for the reverse pedal is stuck from corrosion buildup and sticks when depressed.The complaint was confirmed and the root cause was determined to be a corroded reverse pedal actuator.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTSWITCH,EP-1,PEDAL STYLE,
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7530013
MDR Text Key108941573
Report Number1643264-2018-00390
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010251978
UDI-Public(01)03596010251978
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K892062
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7205396S
Device Catalogue Number7205396
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-