Catalog Number 328418
Device Problems Bent (1059); Difficult to Advance (2920)
Patient Problems Bruise/Contusion (1754); Hemorrhage/Bleeding (1888)
Event Date 05/03/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use a bd insulin syringe with the bd ultra-fine¿ needle was found with multiple issues."the needles have been bent before use, plunger rod is difficult to move, plunger rod is bending during injection, and the consumer experiences bruising and bleeding after injection.".There was no report of exposure, or medical intervention needed.
 
Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch # 7289856.All inspections were performed per the applicable operations qc specifications.There were nine notifications noted that did not pertain to the complaint.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
INSULIN SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7530264
MDR Text Key109012831
Report Number1920898-2018-00347
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382908418034
UDI-Public00382908418034
Combination Product (y/n)N
PMA/PMN Number
K170386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2022
Device Catalogue Number328418
Device Lot Number7289856
Was Device Available for Evaluation? No
Date Manufacturer Received05/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-