• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Device Problem Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2018
Event Type  malfunction  
Event Description
The recipient reportedly experienced loss of lock.External equipment was exchanged, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed the electrode was severed near the electrode ground ring prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The device passed the mechanical tests performed.The failure of this device is attributed to a short from power node to the electrode ground case at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on the electrode ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.
 
Event Description
The recipient reportedly experienced loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7530731
MDR Text Key109029371
Report Number3006556115-2018-00225
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Patient Family Member or Friend
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2016
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age5 YR
-
-