Model Number MMT-7008A
Device Problems False Reading From Device Non-Compliance (1228); Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 05/09/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 400 mg/dl and the sensor glucose was 60 mg/dl at the time of the incident.Customer treated high blood glucose with the insulin pump and declined troubleshooting for it.Sensor glucose and blood glucose was 340 mg/dl.Troubleshooting was done and customer stated that their blood glucose and sensor glucose levels were not in acceptable range.Sensor glucose was lower than blood glucose.Insulin delivery was suspended due to sensor glucose values.Sensor glucose value that triggered the suspend event was 60 mg/dl.Suspend on low limit in sensor settings was 90 mg/dl.Customer advised to monitor and change sensor if issue persists.The device will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7532538
MDR Text Key108848792
Report Number2032227-2018-03705
Device Sequence Number1
Product Code OZO
UDI-Device Identifier20643169541706
UDI-Public(01)20643169541706(17)180523(10)K247P
Combination Product (y/n)N
Reporter Country CodeMX
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/23/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberK247P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/09/2018
Date Device Manufactured11/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-