• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. JELCO® VIAVALVE® SAFETY IV CATHETERS; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. JELCO® VIAVALVE® SAFETY IV CATHETERS; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number 326010
Device Problem Difficult to Advance (2920)
Patient Problem Pain (1994)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
Information was received indicating that this peripheral intravenous catheter would not penetrate the skin without great difficulty.It was noted that the patient experienced pain.No adverse patient effects were reported.
 
Manufacturer Narrative
Ten unused samples of smiths medical jelco® viavalve® safety iv catheters were returned for analysis.The products were examined with testing for force to advance and force to lock.The samples passed all quality requirements.Based on the evidence there were no discrepancies or anomalies found relevant to complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO® VIAVALVE® SAFETY IV CATHETERS
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
201 west queen street
southington CT 06489
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key7535018
MDR Text Key109011319
Report Number3012307300-2018-01899
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier15019517078398
UDI-Public15019517078398
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K160235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date02/28/2020
Device Catalogue Number326010
Device Lot Number3566527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-