Catalog Number 82-3832
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts are being made to obtain the sample.Upon completion of the investigation, a followup report will be submitted.
 
Event Description
As reported by the sales rep, a hakim valve, when primed, was discovered to have a hole, fluid was seen coming out of the valve housing.A replacement valve was used to complete the procedure.There were no reports of delay or patient harm.
 
Manufacturer Narrative
Updated udi : (b)(4).The device has been returned for evaluation.Upon completion of the investigation, a follow-up report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was returned for evaluation.The position of the cam when valve was received was at 120mmh2o.The valve was visually inspected: and confirmed the tear/cut in the silicone housing.A review of manufacturing records found no discrepancies when the device was released to stock.The root cause for the tear/cut in the silicone housing is likely due to the user.As noted in the ifu silicone has a low tear / cut resistance.Previous investigation has concluded that silicone housing tears are not design related, in order for the housing tear to occur the user has to compromise the silicone through a nick or tear in order for the event to occur.Validation testing demonstrates a robust design.Testing has shown that if the silicone housing has been compromised, a housing tear is likely to occur.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMABLE VALVE IN-LINE W/SIPHONGUARD
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
le locle CH 24 00
SZ  CH 2400
MDR Report Key7537759
MDR Text Key109050179
Report Number1226348-2018-10366
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K992173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup,Followup
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Catalogue Number82-3832
Device Lot NumberCVHBDP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Date Manufacturer Received06/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-