CODMAN & SHURTLEFF, INC. / MEDOS S.A. PROGRAMMABLE VALVE IN-LINE W/SIPHONGUARD; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
|
Back to Search Results |
|
Catalog Number 82-3832 |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Attempts are being made to obtain the sample.Upon completion of the investigation, a followup report will be submitted.
|
|
Event Description
|
As reported by the sales rep, a hakim valve, when primed, was discovered to have a hole, fluid was seen coming out of the valve housing.A replacement valve was used to complete the procedure.There were no reports of delay or patient harm.
|
|
Manufacturer Narrative
|
Updated udi : (b)(4).The device has been returned for evaluation.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned for evaluation.The position of the cam when valve was received was at 120mmh2o.The valve was visually inspected: and confirmed the tear/cut in the silicone housing.A review of manufacturing records found no discrepancies when the device was released to stock.The root cause for the tear/cut in the silicone housing is likely due to the user.As noted in the ifu silicone has a low tear / cut resistance.Previous investigation has concluded that silicone housing tears are not design related, in order for the housing tear to occur the user has to compromise the silicone through a nick or tear in order for the event to occur.Validation testing demonstrates a robust design.Testing has shown that if the silicone housing has been compromised, a housing tear is likely to occur.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|