Model Number 400CE
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
Csz received a complaint stating during a case the device set off the current leakage alarm.The case was completed successfully and no patient harm or injury occurred.Csz technical support had the user facility biomed disconnect the heater and run the unit checking for the current leakage.The user facility confirmed current leakage and found the heater to be defective.The facility replaced the heater and the devices functions as designed.
 
Event Description
Csz received a complaint stating during a case the device set off the current leakage alarm.Biomed brought the unit back to the shop and verified the current leakage.The case was completed successfully and no patient harm or injury occurred.The report was filed in our complaint handling system as complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM 400 CE
Type of Device
CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER.
Manufacturer (Section D)
CINCINNATI SUB-ZERO
12011 mosteller road
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller road
cincinnati, OH 45241
MDR Report Key7538718
MDR Text Key109255512
Report Number1516825-2018-00004
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown