ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3262
Device Problems Pacemaker Found in Back-Up Mode (1440); Out-Of-Box Failure (2311)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2018
Event Type  malfunction  
Event Description
It was reported that prior to implant procedure, the pacemaker was noted to be in backup vvi mode in the box. No signs of damage or mishandling were noted. The device was not used.
 
Manufacturer Narrative
The reported field event of backup vvi (bvvi) was confirmed in the laboratory. The device was tested on the bench and analysis of device image revealed a power on reset had occurred. It was also noted that there was code corruption and device id bit corruption. The cause of the bvvi was due to the power on reset but the root cause of the power on reset could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUADRA ALLURE MP RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7538722
MDR Text Key109071956
Report Number2017865-2018-07698
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/31/2019
Device Model NumberPM3262
Device Lot NumberA000051987
Other Device ID Number05414734509091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-